Recall Z-0652-2017 ยท JIX Calibrator, Multi-Analyte Mixture

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA/CRE2), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista System

DateSep 15, 2016
StatusTerminated
ReasonSiemens determined that discrepant low calcium recovery of QC and patient samples can occur with Dimension Vista CHEM 1 CAL lots 5GM081 and 5GM082.
Root causeUnder Investigation by firm
ActionSiemens mailed an Urgent Medical Device recall notification letter on September 19, 2016 to notify affected customers who purchased the product and to make them aware of the potential risk to health and the actions to be taken. For questions regarding tis recall. For questions regarding this recall call 1-800-441-9250.
Quantity9,935 units
DistributionWorldwide Distribution - US (Nationwide) Internationally to Australia, Austria, Bahamas, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Italy, Japan, Netherlands, New Zealand, Norway, Portugal, Qatar, Saudi Arabia, South Korea, Slovakia, Slovenia, Spain, and Swit
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061838DIMENSION VISTA CHEMISTRY 1 CALIBRATOR, MODEL KC110
Dade Behring, Inc., Aug 21, 2006