Recall Z-0315-2017 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-up

DateOct 26, 2016 (initiated Oct 5, 2016)
StatusTerminated
ReasonIntuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in unexpected master movement and potential instrument tip movement under certain circumstances
Root causeSoftware design
ActionUrgent Medical Device Correction letters were sent on October 12, 2016 by traceable method to all affected customers. Letters notify customers of the issue and provided instructions to prevent it from happening until ISI software can install new software upgrade in 1st Q 2017.
Quantity677 devices
DistributionAustralia, Austria, Belgium, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Portugal, Puerto Rico, Qatar, Romania, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United State