Recall Z-0781-2017 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Nuvectra

Algovita Spinal Cord Stimulation System, Percutaneous Leads, 8-Electrode Percutaneous Lead Kit; Algovita Spinal Cord Stimulation System Percutaneous Leads, 12-Electrode Percutaneous Lead Kit; Algovita Spinal Cord Stimulation System Percutaneous Trial Leads, 8-electrode Lead Kit; Algovita Spinal Cord Stimulation System Percutaneous Trial Leads, 12-electrode Lead Kit. Model Number (s) 112x-xx

DateDec 17, 2016 (initiated Oct 11, 2016)
StatusTerminated
ReasonNuvectra is conducting a recall due to two clinical risks that are associated with certain percutaneous leads currently being used in the Algovita Spinal Cord Stimulation System. It has been reported that in some percutaneous leads, the stylet protruded beyond the distal lead tip.
Root causeNonconforming Material/Component
ActionNuvectra field sales associate or contracted distributor were sent a Nuvectra Urgent Field Safety Notice (Field Action Number: 2016-03) on 10/11/2016 via Fed-Ex. The letter provided background information, described the problem, & the product involved in the recall. The letter listed the potential clinical implications, rate of occurrence, recommendations and mitigations and action needed. Provided consignees with notices for them to deliver to physicians as well as Response forms for them and the physician. For questions contact Kathy Jo Fahery +01.763.951.8219.
Quantity406 OUS, 985 US
DistributionUS: Nationwide (CA, TX, KY, FL, IL,PA,OH, GA, NJ, NC, KS, AR, CO, NE, NY, VA, AZ, OK). OUS: Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.