Recall Z-0363-2017 ยท LJE Catheter, Nephrostomy

C.R. Bard, Inc.

X-Force Nephrostomy Balloon Dilation Catheter Kit with Inflation Device, PTFE Sheath, 8 mm x 15 cm, Product Code 996081 The product is a dual lumen catheter with a 24 (8mm) or 30Fr (10mm) balloon mounted on the distal tip. It has a radiopaque tip and radiopaque marker beneath the balloon. It is recommended for use in the dilation of the nephrostomy tact and for placement of the working sheath.

DateOct 12, 2016
StatusTerminated
ReasonContract manufacturer packaged an incorrect sheath with the product. Packaged 30Fr instead of 24Fr.
Root causeUnknown/Undetermined by firm
ActionThe firm, Bard Medical, sent an "URGENT: MEDICAL DEVICE PRODUCT RECALL" letter dated 10/13/2016 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine their inventory and quarantine product subject to recall; do not use or further distribute any affected product; complete and return the attached Recall & Effectiveness Check Form regardless of whether or not you have any of the affect product by fax to 1-770-784-6469 or email a scanned copy to BMD.FieldAction@crbard.com by October 24, 2016, and if you have furth
Quantity192 units
DistributionWorldwide distribution: US distribution to: CA, KS, MO, NY, VA, WA, and to countries of: China, Italy, Sweden and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063632X-FORCE N30 NEPHROSTOMY BALLOON DILATION CATHETER
C.R. Bard, Inc., Jan 5, 2007