ClearanceBrief
Recall Z-0343-2017 ยท DQY Catheter, Percutaneous

Stryker Neurovascular

Guider Softip XF; GUIDER/ST XF/6FR/100 cm, Model number:M003101500 . Cardiology: Boston Scientific Guider Softip XF Guide Catheters are intended to facilitate the placement of interventional devices into the neurovascular system.

DateNov 2, 2016 (initiated Oct 5, 2016)
StatusTerminated
ReasonFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Root causeEmployee error
ActionStryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated October 5, 2016, to all affected customers by Federal Express. Letters advised of the reason for recall and models and devices affected and requested that all affected devices in inventory be returned to Stryker. Customers were advised to circulate the Field Safety Notice internally to all interested/affected parties and to complete the attached customer response form and return the form to their nominated Stryker Representative or to NVFieldActions@stryker.com. For questions regarding this recall call 510-413-2
Quantity1478 devices, all models and lot numbers.
DistributionWorldwide Distribution - US including AK, IL, OH and Internationally to: ARGENTINA AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CHILE CZECH REPUBLIC ECUADOR EGYPT FRANCE GERMANY GREECE HUNGARY INDIA IRAN ISRAEL ITALIA JAPAN MEXICO NETHERLANDS NORWAY PERU PHILIPPINES POLAND RUSSIA SERBIA SINGAPORE SLOVAKI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010853GUIDER SOFTIP GUIDING CATHETER XF 5F
Boston Scientific, Target, Apr 18, 2001
K980453SCHNEIDER GUIDER SOFTIP GUIDING CATHETERS
Boston Scientific Scimed, Inc., Aug 11, 1998