Recall Z-0379-2017 ยท KHP Acid, Lactic, Enzymatic Method

Abbott Laboratories

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.

DateNov 9, 2016 (initiated Oct 11, 2016)
StatusTerminated
ReasonAbbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).
Root causeUnder Investigation by firm
ActionAbbott Laboratories sent an Product Correction Immediate Action Required letter dated October 10, 2016, to all affected customers. Please review this letter with your Medical Director Taking the above information into consideration, you may continue to use ARCHITECT Lactic Acid reagent (LN 9D89-21) Please complete and return the Customer Reply Form If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter Please retain this letter for your laboratory records. If you or any of the health care
Quantity26, 837 kits
DistributionNationwide and Worldwide Distribution.
FDA recordOpen on fda.gov