ClearanceBrief
Recall Z-0774-2017 ยท DYB Introducer, Catheter

Angiodynamics, Inc.

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970505, Catalog No.REF 06597050, 2) 4 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO S PG, Item Number H787065970525, Catalog No.REF 06597052, 3) 4 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO A PG, Item Number H787065970545, Catalog No.REF 0659

DateSep 28, 2016
StatusTerminated
ReasonGreatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
Root causeComponent design/selection
ActionAngiiodynamics sent an Urgent Voluntary Medical Device Recall letter dated September 28, 2016, to all affected customers. Recall notifications were delivered by Federal Express. Customers were instructed to segregate and return all affected devices to AngioDynamics and to complete and return the Reply Verification Tracking Form, provided in the recall notification by fax to 1-800-782-1357. Customers were encouraged to forward a copy of the recall notification to all sites to which have distributed affected product. If affected product is located in your institution, please call AngioDyna
Quantity751 units
DistributionNationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051655ANGIODYNAMICS, INC., MICRO ACCESS KITS
AngioDynamics, Inc., Sep 12, 2005