Recall Z-0117-2017 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
St Jude Medical Inc.
Quadra Assura, Model No. CD3265-40, CD3265-40Q, CD3365-40C, CD3365-40Q. Quadra Assura MP, Model No. CD3269-40, CD3269-40Q, CD3369-40C. Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may also be progra
| Date | Oct 21, 2016 (initiated Oct 10, 2016) |
| Status | Terminated |
| Reason | St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit. |
| Root cause | Device Design |
| Action | St. Jude Medical, Inc., sent a customer letter to all affected customers to inform them that St. Jude Medical recently announced a global medical device advisory for a subset of their company's Fortify, Fortify Assura, Quadra Assura, Unify, Unify Assura and Unify Quadra implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices. The letter informs customers that the cause for Premature Battery Depletion is attributed to the formation of lithium clusters within the battery causing a short circuit. There is a possibility that affected devices ma |
| Quantity | 193,954 units |
| Distribution | Worldwide Distribution - US (nationwide) and OUS |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.