Recall Z-0116-2017 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

St Jude Medical Inc.

Unify, Model No. CD3231-40, CD3231-40Q. Unify Quadra, Model No. CD3249-40, CD3249-40Q. Unify Assura, Model No. CD3257-40, CD3357-40C, CD3357-40Q. Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may als

DateOct 21, 2016 (initiated Oct 10, 2016)
StatusTerminated
ReasonSt Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit.
Root causeDevice Design
ActionSt. Jude Medical, Inc., sent a customer letter to all affected customers to inform them that St. Jude Medical recently announced a global medical device advisory for a subset of their company's Fortify, Fortify Assura, Quadra Assura, Unify, Unify Assura and Unify Quadra implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices. The letter informs customers that the cause for Premature Battery Depletion is attributed to the formation of lithium clusters within the battery causing a short circuit. There is a possibility that affected devices ma
Quantity204,785 units
DistributionWorldwide Distribution - US (nationwide) and OUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.