Recall Z-0115-2017 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)
St Jude Medical Inc.
Fortify VR, Model No. CD1231-40, CD1231-40Q. Fortify ST VR, Model No. CD1241-40, CD1241-40Q. Fortify Assura VR, Model No. CD1257-40, CD1257-40Q, CD1357-40C. Fortify Assura ST VR, Model No. CD1263-40, CD1263-40Q, CD1363-40, CD1363-40Q. Fortify DR, Model No. CD2231-40, CD2231-40Q. Fortify ST DR, Model No. CD2241-40, CD-2241-40Q, CD2263-40, CD2363-40C, CD2363-40Q. Fortify Assura DR, Model No. CD2257-
| Date | Oct 21, 2016 (initiated Oct 10, 2016) |
| Status | Terminated |
| Reason | St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit. |
| Root cause | Device Design |
| Action | St. Jude Medical, Inc., sent a customer letter to all affected customers to inform them that St. Jude Medical recently announced a global medical device advisory for a subset of their company's Fortify, Fortify Assura, Quadra Assura, Unify, Unify Assura and Unify Quadra implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices. The letter informs customers that the cause for Premature Battery Depletion is attributed to the formation of lithium clusters within the battery causing a short circuit. There is a possibility that affected devices ma |
| Quantity | 204,785 units |
| Distribution | Worldwide Distribution - US (nationwide) and OUS |
| FDA record | Open on fda.gov |