ClearanceBrief
Recall Z-0397-2017 ยท MRG System, Test, Vitamin D

Siemens Healthcare Diagnostics, Inc

Siemens ADVIA Centaur Vitamin D Total Assay SMN 10699201 (100 test)

DateOct 5, 2016
StatusTerminated
ReasonVitamin D Total- Change in Correlation Between Serum and Plasma Specimen Tubes- negative bias with plasma vs. serum samples
Root causeDevice Design
ActionSiemens issued an Urgent Medical Device Correction (UMDC) was sent to all affected Siemens Healthcare Diagnostics customers in the United States on October 5, 2016 to be delivered to customers on October 6th, 201 6. An Urgent Field Safety Notice ~UFSN) was sent out for distribution to all customers outside the United States on October 5, 2016. These notices inform the customer of the change in correlation between serum and plasma specimen tube results generated on the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT Vitamin D Total assay. A revised letter was issued on November 7, 2016 by F
Quantity69781 kits
DistributionNationwide Foreign: Afghanistan, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Georgia, Germany, Great Britain, Greece, Guadeloupe, Guatemala,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110586ADVIA CENTAUR VITAMIN D TOTAL ASSAY
Siemens Healthcare Diagnostics, Inc., Oct 14, 2011