Recall Z-0400-2017 ยท FRN Pump, Infusion

Bayer Healthcare

PET Infusion Pump

DateNov 11, 2016 (initiated Sep 30, 2016)
StatusTerminated
ReasonThe Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one year past manufacturing date.
Root causeProcess control
ActionBayer mailed an Urgent Medical Device Customer Notification letter to the affected consignees on September 30, 2016 asking that they review their current SAS inventory and determine if you have any product that is more than one year past its manufacturing date, Quarantine unused units and call Bayer Customer Care line to receive a Return Goods Authorization (RGA) number for the product's return.
Quantity165,127 (79,714 units in U.S.)
DistributionDistributed Nationwide and internationally to : Taiwan, Ukraine, Turkey, Thailand, Slovania, Singapore, Sweden, Saudi Arabia, Serbia, Romania, Qatar, Portugal, Poland, Panama, Oman, Norway, The Netherlands, Mexico, Malta, Morraco, Libyan, Lithuania, Lebanon, Kuwait, Italy, Israel, Hungray, United Ki
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080297MEDRAD INTEGO PET INFUSION SYSTEM
Medrad, Inc., Jun 17, 2008
K100798MEDRAD INTEGO PET INFUSION SYSTEM, SOURCE ADMINISTRATION SET, PATIENT ADMINSTRATIN SET MODEL INT SYS 100, INT CSS, INT C
Medrad, Inc., Jun 23, 2010