ClearanceBrief
Recall Z-0875-2017 ยท JAK System, X-Ray, Tomography, Computed

Neusoft Medical Systems Co., Ltd.

NeuViz 64 Multi-slice CT Scanner Systems

DateAug 31, 2016
StatusTerminated
ReasonThe Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error before scanning can resume.
Root causeSoftware design
ActionNeusoft Medical Systems Co. Ltd. planned action to bring into compliance. 1.Affected Customers will be notified by letter to provide them a description of the affected device, instructions for using the device prior to correction, hazards associated with the defect, and a statement of corrective actions to be taken at no cost to the owner. 2.Field Service Engineers will visit owner sites to install a software update which resolves the defect. 3.Corrections will be made at no cost to the owner 4.Corrections will be completed by May 11, 2017. CDRH approves the CAP subject to the following c
Quantity23 units
DistributionUS Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121792NEUVIZ 64 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Nov 16, 2012