Recall Z-0395-2017 ยท LCP Assay, Glycosylated Hemoglobin

Roche Diagnostics Operations, Inc.

Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190

DateNov 11, 2016 (initiated Oct 4, 2016)
StatusTerminated
ReasonRoche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increase
Root causeStorage
ActionOn 10/4/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Direct Consignees and Distributors via courier. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Direct Consignee Actions Required " Follow the instructions in the Customer Actions for Prevention/Detection of the Issue section of this Urgent Medical Device Correction (UMDC). " Complete the attached fax form (7184-00-1016) and fax it to 1-844-294-7181. " File this U
Quantity4976
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072714TINA-QUANT HEMOGLOBIN A1C GEN.2 TEST SYSTEM
Roche Diagnostics Corp., Apr 18, 2008