ClearanceBrief
Recall Z-0107-2017 ยท JQP Calculator/Data Processing Module, For Clinical Use

Roche Diagnostics Operations, Inc.

cobas p 512 Pre-Analytical Instrument calculator/data processing module, for clinical use Product Usage: The cobas 6800/8800 systems support an automated and integrated workflow to run Polymerase Chain Reaction (PCR) based Nucleic Acid Testing (NAT) to be applied in diagnostic laboratories.

DateJul 15, 2016
StatusTerminated
ReasonRoche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Root causeDevice Design
ActionOn 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier
Quantity9014 in total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.