ClearanceBrief
Recall Z-0576-2017 ยท GBZ Catheter, Cholangiography

Atrium Medical Corporation

Atrium Medical 8 Fr Trocar catheter, Sterile Model Number: 8408 Intended to facilitate the evacuation of air and/or fluid from the chest cavity or mediastinum.

DateOct 26, 2016
StatusTerminated
ReasonTrocar Catheters are packaged with a label containing an icon that shows more holes (eyelets) than the product actually contains.
Root causeLabeling design
ActionAtrium Medical issued an "Urgent Medical Device Recall (Removal)" communication to customers on 10/26/16 via FedEx 2 day delivery. The recall identifies the problem, product, and risk factors. In addition, it provides instructions on how to handle product that may be affected. Accounts are to read the letter completely and share with all users and staff. They have been requested to complete the enclosed Recall Reply Form, acknowledging receipt of the notice; and, remove and quarantine any affected product for return. Additionally, obtain product from an alternate supplier because replaceme
Quantity1617 cases (10/cs)
DistributionWorldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Costa Rica, Dominican Republic, Ecuador, Finland, France, Germany, Honduras, Hong Kong, Ireland, Jamaica, Kuwait, Mexico, Netherlands, New Zealand, Panama, Peru, Portugal, Puerto Rico, Qatar, Reunion, Romania, Saudi Arabia, South A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K831931ATRIUM THORACIC CATHETER
Atrium Medical Corp., Oct 4, 1983