Recall Z-0636-2017 ยท FDF Colonoscope And Accessories, Flexible/Rigid

Pentax of America Inc

Pentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

DateNov 19, 2016 (initiated Aug 15, 2016)
StatusTerminated
ReasonPentax is initiating the product correction of several model numbers of Pentax's Colonoscope that were shipped between the clearance date of April 2014 and September 2015 because they contain the operation and reprocessing IFU's that were in use prior to the 2014 clearances and therefore did not include the cleared operation IFU, the updated reprocessing IFU or the dedicated reprocessing adaptor (for applicable models).
Root causeOther
ActionPentax sent as "Urgent Field Correction" and Field Correction Response Form dated August 15, 2016 to their affected customers.
Quantity5,090 (US) and 14,209 (OUS)
DistributionUS Nationwide and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131855PENTAX VIDEO COLONOSCOPES (EC FAMILY)
Pentax Medical Company, Apr 9, 2014