Recall Z-0372-2017 ยท JEB Table, Surgical With Orthopedic Accessories, Manual

Schaerer Medical USA Inc

Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are part number 85118. Product Usage: The cylinders are used to lock the surgical table in place.

DateSep 27, 2016
StatusTerminated
ReasonThe floor locking cylinders are contained in the base of the Schaerer 7300 model surgical tables. The cylinders were manufactured with the incorrect "O" ring which may cause the cylinder to leak.
Root causeNonconforming Material/Component
ActionSchaerer Medical sent a Field action letters dated September 27, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The customer is required to complete and return a response form with the number of affected tables and cylinders. All affected cylinders will be replaced in the field and returned to the manufacturer. For questions Quality Assurance at 513-272 -9116.
Quantity10 replacement cylinders.
DistributionWorldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.