Recall Z-0371-2017 ยท JEB Table, Surgical With Orthopedic Accessories, Manual

Schaerer Medical USA Inc

The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable. Product Usage: The cylinders are used to lock the surgical table in place.

DateSep 27, 2016
StatusTerminated
ReasonThe floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause the cylinder to leak.
Root causeNonconforming Material/Component
ActionSchaerer Medical sent a Field action letters dated September 27, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The customer is required to complete and return a response form with the number of affected tables and cylinders. All affected cylinders will be replaced in the field and returned to the manufacturer. For questions Quality Assurance at 513-272 -9116.
Quantity4 surgical tables with defective floor locking cylinders were distributed. Each table contains 3 cylinders for a total
DistributionWorldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed