Recall Z-0725-2017 ยท JGJ Photometric Method, Magnesium

Randox Laboratories, Limited

Magnesium on RX Imola analyser IVD

DateApr 14, 2016
StatusTerminated
ReasonAccording to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.
Root causeDevice Design
ActionRandox sent an Urgent Field Safety Notice to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to review and update Rx Imola running order where required. Rework any remaining stock with updated IFU and Important Notice. Complete and return the vigilance response section of the form to technicalservikces@randox.com.
Quantity15
DistributionNationwide Distribution The product was only distributed to US Consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052914RX IMOLA
Randox Laboratories, Ltd., Mar 31, 2006