Recall Z-0402-2017 ยท JIY Photometric Method, Iron (Non-Heme)

Siemens Healthcare Diagnostics, Inc.

Dimension IRON Flex reagent cartridges

DateAug 25, 2016
StatusTerminated
ReasonSiemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).
Root causeError in labeling
ActionSiemens mailed an Urgent Medical Device Correction letter to all customers who received the Dimension and Dimension Vista IRON Flex reagent with an update to the Limitation of Procedure section for the Dimension and Dimension Visa IRON assays Instructions For Use (IFU) to indicate that: patients treated with metal-binding drugs (e.g. deferoxamine) may have depressed iron values, as chelated iron may not properly react in the iron assay. Customers were told that the information provided in the letter, related to deferoxamine, supersedes the information related to deferoxamine int he current
Quantity38,230 cartridges
DistributionDistributed Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060264DIMENSION IRON FLEX REAGENT CARTRIDGE (IRON-DF85)
Dade Behring, Inc., Mar 8, 2006