Recall Z-0745-2017 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Becton Dickinson & Company

BD 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from patients.

DateAug 8, 2016
StatusOpen, Classified
ReasonBD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon activation of the safety feature, which retracts the needle.
Root causeDevice Design
ActionThe firm, BD, sent a "URGENT PRODUCT RECALL" letter dated 8/8/2016 to its customer with a return response form via UPS 2nd Day mail on 8/8/2016.The letter described the product, problem and actions to be taken. The customers were instructed to immediately review your inventory for the specific lot numbers, quarantine the product and discontinue the shipment of the affected product; complete and return the Recall Response Form by fax to BD at 1-866-913-0661 or email to bd6835@stericycle.com (even if you do not have any product); return all affected products with completed response form per pac
Quantity1,788,150 units
DistributionWorldwide distribution- US (nationwide) and countries of: Canada, Belgium, Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030573BD VACUTAINER PUSH BUTTON BLOOD COLLECTION SET, MODELS 367338, 367344, 367336, 367342, 367335, 367341
Bd Vacutainer Systems Preanalyical Solutions, May 8, 2003