ClearanceBrief
Recall Z-0083-2017 ยท FZX Guide, Surgical, Instrument

Zimmer Biomet, Inc.

Phoenix Recon Soft Tissue Sleeve Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage: The Phoenix Femoral Nail System is to be implanted into the femur for alignment, stabilization and fixation of fractures caused by trauma or disease, and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity,

DateOct 14, 2016 (initiated Aug 9, 2016)
StatusTerminated
ReasonOld revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new revision targeting arm. Risks of encountering the affected product include: the correct instrument may not be readily available for the procedure causing the surgeon to complete the procedure by hand; or a delay in surgery greater than 30 minutes may occur while another Recon Targeting Arm or Recon Soft tissue Sleeve is located or while the sur
Root causeFinished device change control
ActionOn 8/9/2016, URGENT MEDICAL DEVICE RECALL REMOVAL notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions or concerns call the customer call center at 1-800-348-9500 ext 1251 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, your questions may be sent by email to CPWARFieldAction@zimmerbiomet.com.
Quantity31
DistributionUS Nationwide Distribution in the states of: MI, UT, SD, OH, WA, KS, MO, NJ, GA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.