Recall Z-0325-2017 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Encore Medical, Lp

Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.

DateSep 19, 2016
StatusTerminated
ReasonIt has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.
Root causeUnder Investigation by firm
ActionThe recalling firm sent a recall notification letter to affected consignees on 9/19/16. The field safety notice provided a newly revised Surgical Technique to educate users on proper use of the device.
Quantity436 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160342EMPOWR PS Knee System
Encore Medical L.P., May 19, 2016