Recall Z-0326-2017 ยท FSA Lift, Patient, Non-Ac-Powered

Hill-Rom, Inc.

Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.

DateOct 31, 2016 (initiated Aug 9, 2016)
StatusTerminated
ReasonCeiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
Root causeMixed-up of materials/components
ActionOn 8/9/2016, URGENT: MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this Field Safety Correction, please contact Hill-Rom Technical Support at 1-800-445-3720, your distributor or your Hill-Rom representative.
Quantity188
DistributionWA NC OH PA MI Australia Sweden
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.