Recall Z-0609-2017 ยท HWC Screw, Fixation, Bone

Trilliant Surgical Ltd.

Model 208-70-110, Long Thread Tiger Large Cannulated Headless Screw, 7 x 110 mm labeled as "7.0 x 110mm LTTHCS REF 208-70-110 QTY 30 LOT TSL003721 2016-03-29 Trilliant Surgical, LTD"

DateNov 17, 2016 (initiated Sep 2, 2016)
StatusTerminated
ReasonScrews were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.
Root causeNonconforming Material/Component
ActionThe firm sent out written notifcation of the recall on 9/2/16; the letter instructed the consignees to discontinue use of the affected lot, quarantine, and return product of the affected lot in accordance with an included acknowledgement sheet.
Quantity11
DistributionCA, AZ, WI, FL, NJ, OH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112737TIGER HEADLESS CANNULATED SCREWS
Trilliant Surgical, Ltd., Oct 20, 2011