Recall Z-0004-2017 ยท MKJ Automated External Defibrillators (Non-Wearable)

Leonhard Lang Medizintechnik GmbH

Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.

DateOct 14, 2016 (initiated Sep 1, 2016)
StatusTerminated
ReasonThere is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
Root causeDevice Design
ActionConsignees was sent via e-mail a Leonhard Lang "Important Safety Notice" dated September 1, 2016. The letter described the product being recalled, Description of the defect, Actions and time frame of the recall, & Compensation for the recalled electrodes. Advised consignees to inform all users within their organizations of the recall; and, secure and destroy the unused electrodes in their inventory. Confirmation of destruction can be done by completing and returning the "Confirmation of Destruction / Consumption" form by October 14, 2016. If the product was further distributed, they were t
Quantity11,110 ( US 8,040; OUS 3,070)
DistributionWorldwide Distribution - US, to the state of Florida; and, the countries of South Africa, Germany, Slovenia, France, United Arab Emirates, Great Britain, Italy, Israel, Serbia, Thailand, Thailand, South Africa, Lebanon, and Poland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142803Skintact, various other tradenames
Leonhard Lang GmbH, Dec 2, 2014