ClearanceBrief
Recall Z-0010-2017 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems

Ingenuity Core Model Number 728321 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

DateAug 24, 2016
StatusTerminated
ReasonThe infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults t
Root causeSoftware design
ActionPhilips sent an Urgent Field Safety Notice (FSN) CLE 16-045 letter dated August 26, 2016, via a certified letter to customers. The letter identified the affected product, problem, actions to be taken by customer/user and actions planned by Philips to correct the problem. The letter informed customers a Philips field service representative will install a software update on the affected systems. For questions contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompt
Quantity358
DistributionWorldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bang
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003