Recall Z-2913-2016 ยท JFS Coulometric, Chloride

Siemens Healthcare Diagnostics Inc

Siemens RAPIDPoint 405 Measurement Cartridge Siemens SMN numbers: 10844812, 1028322, 10313971 , 10310469

DateSep 30, 2016 (initiated Sep 2, 2016)
StatusTerminated
ReasonError in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge
Root causeUnder Investigation by firm
ActionSiemens Healthcare issued An Urgent Field Safety Notice via e-mail to affected Siemens Healthcare Diagnostics Regional offices on September 2, 2016 both in the United States and outside of the United States, for communication with affected customers e: the bias of the chloride values. US accounts sent hard copy via Federal Express and requested to: Complete and return the Field Correction Effectiveness Check form attached to this letter within 7 days. Review this letter with your Medical Director. Questions:Contact your Siemens Customer Care Center or your local Siemens technical support repr
Quantity6638 cartridges
DistributionNationwide: Foreign: Canada, Bahamas,Germany, Japan, Mexico.Belgium, Italy,Spain, Korea, France, Austria, Korea, France, Poland, Australia, china, AE, Bangladesh, Columbia, Czechkoslovakia, Bosnia
FDA recordOpen on fda.gov