ClearanceBrief
Recall Z-0357-2017 ยท FPA Set, Administration, Intravascular

Unomedical As

AC TenderLink 13/30 10 pcs; AC TenderLink I 13/60 10 pcs; AC TenderLink I 13/80 10 pcs; AC TenderLink I 17/110 10 pcs; AC TenderLink I 17/30 10 pcs; AC TenderLink I 17/60 10 pcs; AC TenderLink I 17/80 10 pcs; AC TenderLink I 13/110 10 pcs The products are used in combination with insulin infusion pumps for continuous subcutaneous infusion of insulin

DateNov 7, 2014
StatusTerminated
ReasonTubing detachment
Root causeDevice Design
ActionConvaTec sent an Urgent Medical Device Correction letter to all affected consignees on November 7, 2014. The letter notified the consignees of what action they need to take if they experience a detached tubing. You can continue to use your infusion sets by following the advice below. 1. When changing your infusion set, closely follow the instructions for use included in the product box. Check the tubing at the connect/disconnect location identified in the drawings above to make sure it is not loose. 2. As always, it is essential to monitor your blood sugar levels frequently using your bloo
Quantity653,280 units
DistributionWorldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120872ASANTE CONSET AND ASANTE COMFORT SUBCUTANEOUS INFUSION SETS FOR USE WITH ASANTE PEARL INFUSION PUMPS
Unomedical A/S Infusion Devices, Apr 20, 2012
K972135MAERSK MEDICAL PURELINE COMFORT SUBCUTANEOUS INFUSION SET
P/L Biomedical, Aug 26, 1997