Recall Z-0319-2017 ยท LKQ Antibody Igm,If, Cytomegalovirus Virus

Siemens Healthcare Diagnostics, Inc.

IMMULITE/ IMMULITE 1000 Systems CMM CMV IgM Assay, REF/Catalog Number LKCM1(D) - 100 tests (US), Siemens Material Number (SMN) 10370301 (US), Unique Device Identification Number (UDI) 00630414985718 (US); and REF/Catalog Number LKCM1 - 100 tests (OUS); SMN 10381296 (OUS), UDI 00630414963747 (OUS); IVD --- Made in UK Siemens Healthcare Diagnostics Products Ltd., Glyn Rhonwy, Llanberis, Gwynedd, LL

DateAug 30, 2016
StatusTerminated
ReasonMultiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above.
Root causeUnder Investigation by firm
ActionThe firm, Siemens, issued an "Urgent Medical Device Recall" Letter (IMC16-22.A.US, dated 8/30/2016) in the US and an "Urgent Field Safety Notice" (IMC16-22.A.OUS, dated August 2016) Outside the US (OUS) starting on 8/30/2016 The notices instructed customers to discontinue use of and discard the affected kit lots listed in the customer notice; review your inventory of the products to determine your laboratory's replacement needs; review the UMDR and or the UFSN with their Medical Director, In the US, the method of distribution is FedEx and Outside the US the distribution is determined at the c
QuantityUS: 5 kits; Foreign: 372 kits
DistributionWorldwide distribution- US (Nationwide) and countries of: United Arab Emirates, Afghanistan, Angola, Argentina, Australia, Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Belarus, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Dominican Republic, Algeria,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100433IMMULITE 2000 CMV IGM, IMMULITE CMV IGM CONTROLS, AND IMMULITE 200 CMV IGM ADJUSTOR MODEL L2KCM2, L2KCM6, LCMC1, LCMC2,
Siemens Healthcare Diagnostics, May 13, 2010