Recall Z-2897-2016 ยท EIH Powder, Porcelain

Prismatik Dentalcraft, Inc

BruxZir Shaded, Model No. 70-1167-NOW0001, 70-1167-NOW0002, 70-1167-NOW0003, 70-1167-NOW0005, 70-1167-NOW0006 The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

DateAug 15, 2016
StatusTerminated
ReasonPrismatik Dentalcraft is recalling BruxZir Shaded because the new design was manufactured at risk and were to be quarantined and held until the validation was completed. Prismatik failed to properly identify the lots as quarantined product.
Root causeRelease of Material/Component prior to receiving test results
ActionPrismatik DentalCraft, Inc. sent an Urgent Medical Device Recall letter dated August 15, 2016, to all their customers to inform them that Prismatik Dentalcraft is recalling BruxZir NOW product because the new design was manufactured at risk and was supposed to be quarantined and held until validation was completed. Prismatik failed to properly identify the implicated lots as quarantined product. Customers are informed that the BruxZir Now with the thicker sprue design was packaged and shipped to customers in error. Customers are informed of that there is no risk to the patient and of the ac
Quantity150 units
DistributionNationwide distribution to SC, MN, NC, IN, AZ, OR, WA, CA, WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130924BRUXZIR SHADED
Prismatik Dentalcraft, Inc., May 16, 2013