Recall Z-2849-2016 ยท HRS Plate, Fixation, Bone

Trimed Inc

TriMed Semi-Tubular Plate, 10-Hole, Model No. SMTP-10 Contoured plate with offset screw holes. The plate is 103 mm in length, 1.3 mm thick and 9.6 mm wide. The part is made of surgical steel conformance with ASTM F138 or F139. The plate is used to fixation of long bone fractures. The TriMed Bone Plates are intended for use in the fixation of fractures to the Tibia, Fibula, Ulna, Radius and the Hu

DateAug 29, 2016
StatusTerminated
ReasonTriMed is recalling the Semi-Tubular Plates with 10 holes (SMTP-10) because a non-locking screw passed through a hole in a SMTP-10 plate being implanted.
Root causeDevice Design
ActionThe firm, TriMed, sent a "Medical Device Recall Notification" Letter dated 8/29/16 to customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and quarantine the product subject to this recall; notify your customers at once of this product recall; discontinue use; return product, and complete and return the enclosed MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form as soon as possible by Fax to: 1-661-254-8485 or email to: davidmedoff@trimedortho.com. Customers with questions are in
Quantity30 units
DistributionWorldwide Distribution-US (nationwide) including states of: Alabama, Florida, Indiana, Connecticut, California, Wisconsin, Missouri, Florida, Hawaii; and countries of: Sweden and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060041TRIMED BONE PLATES
TriMed, Inc., Feb 7, 2006