Recall Z-2885-2016 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Spacelabs Healthcare Inc

Spacelabs Healthcare Xhibit Central Station, Model 96102, Version 1.1.2. Xhibit Telemetry Receivers (XTR) must be connected to an Xhibit Central Station for the display of patient vital signs.

DateSep 26, 2016 (initiated Aug 25, 2016)
StatusTerminated
ReasonThe firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
Root causeSoftware design
ActionSpacelabs sent a letter dated August 25, 2016 to affected US consignees informing them of the going offline or locking up issue when Xhibit Central Station is used as part of an Xhibit Telemetry System. On September 1, 2016, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs sent the second Urgent - Medical Device Correction letter, dated September 15, 2016, to affected US consignees informing them that they have received reports of certain short-duration asystole alarms generated by the Xhibit Telemetry Receiv
Quantity798 units total (488 units in the US and 310 units outside US)
DistributionWorldwide distribution. US nationwide including Puerto Rico, Australia, Canada, China, France, Germany, Ireland, Italy, Netherlands, Philippines, Poland, Romania, Switzerland, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122146XHIBIT CENTRAL STATION, MODEL 96102
Spacelabs Healthcare, Apr 22, 2013