Recall Z-2885-2016 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Spacelabs Healthcare Inc
Spacelabs Healthcare Xhibit Central Station, Model 96102, Version 1.1.2. Xhibit Telemetry Receivers (XTR) must be connected to an Xhibit Central Station for the display of patient vital signs.
| Date | Sep 26, 2016 (initiated Aug 25, 2016) |
| Status | Terminated |
| Reason | The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital. |
| Root cause | Software design |
| Action | Spacelabs sent a letter dated August 25, 2016 to affected US consignees informing them of the going offline or locking up issue when Xhibit Central Station is used as part of an Xhibit Telemetry System. On September 1, 2016, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs sent the second Urgent - Medical Device Correction letter, dated September 15, 2016, to affected US consignees informing them that they have received reports of certain short-duration asystole alarms generated by the Xhibit Telemetry Receiv |
| Quantity | 798 units total (488 units in the US and 310 units outside US) |
| Distribution | Worldwide distribution. US nationwide including Puerto Rico, Australia, Canada, China, France, Germany, Ireland, Italy, Netherlands, Philippines, Poland, Romania, Switzerland, and United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K122146 | XHIBIT CENTRAL STATION, MODEL 96102 Spacelabs Healthcare, Apr 22, 2013 |