ClearanceBrief
Recall Z-0683-2017 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

MicroPort Orthopedics Inc.

Lineage Liners: REF 70102846, REF 70202852, REF 70202852, REF 70302858, REF 70303258, REF 70402864, REF 70403264, REF 70403664 For use with the conserve metal head

DateDec 1, 2016 (initiated Jun 21, 2016)
StatusTerminated
ReasonRevision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Root causeDevice Design
ActionMicroPort sent an Urgent Field Safety Notice dated June 22, 2016, to all affected customers notifying them of the change in the instructions for use on July 7th, 2016 by email. The firm also notified US hospitals where the implantation surgeries occurred within the last twelve (12) months by letter on July 13th, 2016 delivered by FedEx Priority. Customers were asked to complete and return the enclosed form even if they do not have any affected product to return by fax to 1-901-451-6032 or PostMarket@ortho.microport.com. For questions regarding this recall call 901-867-4771.
Quantity5,009 units
DistributionWorldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100866PROFEMUR HIP SYSTEM MODULAR NECKS
Wrightmedicaltechnologyinc, Apr 28, 2010