Recall Z-0321-2017 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Mindray DS USA, Inc. dba Mindray North America

TD60 display Telemetry Transmitter, SHENZHEN Mindray Bio-Medical Electronics Co., LTD, Mindray Building, Keji 12th Road South, Hi-Tech Industrial Partk, Nanshan, Shenzhen, 518057 P.R. China

DateSep 8, 2016
StatusTerminated
ReasonMindray has become aware that the TD60 Display Telemetry Transmitter used with the BeneVision Central Station may crack on the bezel surrounding the display and also the transmitter may not power on or will power on but display an error message.
Root causeDevice Design
ActionMindray sent a "Medical Device Action" letter dated September 8, 2016 via UPS to their affected customers.
Quantity1,278 units
DistributionUS and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150632Hypervisor IX Monitoring System
Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Jul 6, 2015