Recall Z-2859-2016 ยท HQL Intraocular Lens

Abbott Medical Optics Inc. (AMO)

TECNIS OptiBlue 1-Piece Intraocular Lens (IOL), Model No. ZCB00V The product is indicated for implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction. The lens is intended to be placed in the capsular bag.

DateSep 21, 2016 (initiated Aug 26, 2016)
StatusTerminated
ReasonAbbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
Root causeProcess control
ActionAbbott Medical Optics, Inc. sent an Urgent Field Safety Notice to all affected customers to inform them that Abbott Medical Optics (AMO) is recalling 737 IOLs due to detection of an inspection equipment malfunction that resulted in a total of 737 distributed lenses not being fully checked prior to release. The letter informs customers that this malfunction may result in the release of mislabeled IOLs. The letter informs the customers of the actions to be taken. Customers are instructed to complete Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to RegCompli
Quantity131 units
DistributionWorldwide Distribution - US nationwide, PR, and Internationally to Canada, Latin America, Europe, Africa, Middle East, and Asia Pacific.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.