ClearanceBrief
Recall Z-2918-2016 ยท HTN Washer, Bolt Nut

Arthrex, Inc.

Arthrex Suture Washer

DateAug 12, 2016
StatusTerminated
ReasonThe device was assembled incorrectly. Suture and washer were not assembled; components were placed loose in the package.
Root causePackaging process control
ActionConsignees were notified 8/12/2016 of the potentially affected devices; Product/Batch Numbers: AR-7000-818T Batch #10028617. For Healthcare Professionals: immediately discontinue use of all affected products with stated numbers in their possession. Call (855) 552-5056 or email returns@arthrex.com with the product number, lot, and quantities so they may ship replacement products and expedite return of recalled product. To please complete the "Recall Acknowledgement of Receipt: postal card and return to Arthrex, Inc. Distributing Agencies: Immediately identify and hold all stated produc
Quantity45 devices
DistributionOR, NY, MN, KY, CA, ID, CO, MA, and FL. Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043145BI-CORTICAL BIO-POST AND WASHER, MODEL AR-1367B
Arthrex, Inc., Feb 18, 2005