Recall Z-2854-2016 ยท KRA Catheter, Continuous Flush

EKOS Corporation

EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter

DateAug 9, 2016
StatusTerminated
ReasonThe recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant parameters, which may result in a small blister in the epoxy over transducers in the treatment zone. It is not known at the time the recall was initiated whether the non-compliant test parameters, or presence of the blisters, adversely affect operation of the device.
Root causeStorage
ActionAll affected consignees were contacted by e-mail on August 9, 2016, and asked to sequester implicated devices.
Quantity10 units
DistributionThree international consignees - Germany, New Zealand and Taiwan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081467EKOSONIC ENDOVASCULAR SYSTEM WITH RAPID PULSE MODULATION
Ekos Corp., Jun 26, 2008
K111705EKOSONIC ENDOVASCULAR SYSTEM
Ekos Corporation, Aug 3, 2011
K140151EKOSONIC MACH 4 ENDOVASCULAR DEVICE/EKOSONIC CONTROL UNIT
Btg International, Inc., May 21, 2014