Recall Z-2776-2016 ยท DQK Computer, Diagnostic, Programmable

Biosense Webster, Inc.

CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO 3 EP Navigation System is intended to be used in catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and catheter location during the procedure.

DateSep 9, 2016 (initiated Aug 15, 2016)
StatusTerminated
ReasonBiosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).
Root causeUnder Investigation by firm
ActionBiosense Webster sent a Customer Notification letter dated August 23, 2016, to all affected customers to inform them that Biosense Webster, Inc. recently became aware that some systems were improperly transported and installed at hospitals in China by Biosense Webster, Inc. Clinical Specialists. The letter informs the customers that according to the CARTO 3 System Instructions for Use (IFU), only the Biosense Webster, Inc. Technical Services team members are authorized to properly install the systems. Customers are informed that each impacted CARTO 3 System must be fully tested and approved b
Quantity338 units
DistributionInternational Distribution to China, Hong Kong and Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112007CARTO 3 EP NAVIGATION SYSTEM, VERSION 2.2
Biosense Webster, Inc., Oct 5, 2011
K132782CARTO3 EP NAVIGATION SYSTEM, VERSION 3.2
Biosense Webster, Inc., Jan 24, 2014
K133916CARTO 3 EP NAVIGATION SYSTEM, VERSION 4.2
Biosense Webster, Inc., Jul 1, 2014