Recall Z-0006-2017 ยท DSQ Ventricular (Assist) Bypass

HeartWare, Inc

HeartWare Ventricular Assist System (HVAD), US product: Model No.: 1103; and International Product: Model No.: 1104XX. Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. Designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.

DateAug 17, 2016
StatusOpen, Classified
ReasonForeign material found within the driveline connector of the HVAD Pump that could cause increased impedance resulting in electrical faults.
Root causeComponent design/selection
ActionHeartWare notified their customers with an Urgent Medical Device Recall notice on 8/17/2016 of the HVAD Pump, listing the US Product: Model No. 1103 and the International Product: Model No.: 1104. The letter discussed the reason for the recall as well as the risk to health. Customers were to identify affected product in their inventory; and, complete and return the acknowledgement form. Acknowledgement should be provided no later than 30 days from date of the letter either to the HeartWare representative; or, via e-mail to FSCA @Heartware.com; or, FAX the signed form to (305) 364-2665. Af
Quantity344 pumps
DistributionWorldwide Distribution - US, in the states of AL, AR, AZ, CA, DC, FL, IL, IN, KY, LA, MA, MI, MN, MO, NC, NE, NJ, NY, OH, OK, PA, SC, TX, UT, VA, and WI; and, the countries of Argentina, Belarus, Bosnia, Herzegovina, Brazil, Canada, Chile, Colombia, Hong Kong, India, Israel, Kazakhstan, Kuwait, Leba
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.