Recall Z-2768-2016 ยท DXT Injector And Syringe, Angiographic

Bayer Healthcare

Certegra Workstation used in conjunction with Medrad Stellant CT Injection System Catalog Number # CWKS DISPLAY Serial Numbers The Medrad Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications

DateAug 2, 2016
StatusTerminated
ReasonCertegraWorkstation when used in conjunction with the Medrad Stellant CT Injection System may exhibit thermal events, such as: smoke and sparks of fire.
Root causeUnder Investigation by firm
ActionBayer Healthcare mailed an Urgent Field Safety Notification letter on August 03, 2016 to customers who were affected by the recall informing them of the issue and what actions to take if they were impacted. Customers were advised to use the enclosed Interim Instructions for Use for further information. For questions customers should call 1-877-229-3767.
Quantity165 systems
DistributionWorldwide Distribution - US (nationwide) and Internationally to United Arab Emirates, Australia, Canada, Germany, Spain, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082905MEDRAD STELLANT CT INJECTOR SYSTEM WITH P3T CARDIAC
Medrad, Inc., Jan 12, 2009