Recall Z-2691-2016 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

HexaPOD evo RT System Product Usage: The intended use of the device is the control of accurate patient positioning with assistance of a 3D Tracking System in a radiotherapy environment.

DateAug 30, 2016 (initiated Aug 18, 2016)
StatusTerminated
ReasonPotentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position.
Root causeNonconforming Material/Component
ActionElekta sent an Important User Notice 618-03-303-025 to affected customers on 8/18/2016. The notice identified the affected product, problem, clinical impact, solution and actions to be taken. The Notice reminds users that the Interlock Check is not possible during 3D Workflow as iGUIDE is not in use. The notice reminds users that if iGUIDE is not being used - (A) Ensure the HexaPOD is at its pre-defined 3D position before treatment, and (B) Ensure the External Inhibit LED at the EnableSwitch Board is off. For questions contact your local Elekta office.
Quantity13
DistributionWorldwide Distribution - US Nationwide in the states of LA, PA, WA, and the countries of: Australia, Denmark, France, India, Italy, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072079IGUIDE SYSTEM