Recall Z-2712-2016 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco RTP System Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

DateAug 18, 2016
StatusTerminated
ReasonIn a specific workflow where contours are edited (enlarged, moved, copied, etc.) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present which would not be the intent of the planner.
Root causeUnknown/Undetermined by firm
ActionImportant User Notice 382-03-MON-003 was sent to customers on August 18, 2016. The notice informs users of the specific product and version numbers affected by the issue, and any work arounds that can be used to avoid the issue. The letter also states that the problem will be in a future Monaco release. The customers were instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta as soon as possible but no longer than 30 days from receipt. Questions can be directed to the local Elekta office.
Quantity3,012 units
DistributionWorldwide Distribution -- USA, to the states of AK, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, MT, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, Puerto Rico, Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Bahrain, Bangladesh, B
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110730MONACO RTP SYSTEM
Computerized Medical Systems, Inc., Jun 24, 2011