Recall Z-2899-2016 ยท DDR Myoglobin, Antigen, Antiserum, Control

Siemens Healthcare Diagnostics, Inc

Siemens ADVIA Centaur Calibrator U, For in vitro diagnostic use in calibrating the following assays using ADVIA Centaur"' systems: Myoglobin cTnl Cat No. ADVIA Centaur Calibrator U (2 Pack) - 03684480 (SMN 1 0309996)

DateAug 8, 2016
StatusTerminated
ReasonFailure of analytical sensitivity with Calibrator U kit lots ending in 63 and 64 and a drift from the internal standardization
Root causeNonconforming Material/Component
ActionSiemens Healthcare Diagnostics issued an Urgent Medical Device Correction (UMDC) to customers in the US on August 8, 2016, to be delivered to customers on August 9, 2016. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on August 8, 2016. The communications notify customers of the failure of analytical sensitivity with Calibrator U kit lots ending in 63 and 64 and a drift from the internal standardization. The communications also inform customers that analytical sensitivity and alignment to the internal standardization will be re
Quantity2,248 kits
DistributionWorldwide Distribution - Nationwide Distribution including the countries of: Argentina Australia Brazil Canada Chile China Ecuador India Japan Kore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122599ADVIA CHEMISTRY MYOGLOBIN REAGENT, ADVIA CHEMISTRY MYOGLBIN CALIBRATOR
Siemens Healthcare Diagnostics, Inc., Oct 3, 2012