Recall Z-2686-2016 ยท LMB Device, Digital Image Storage, Radiological
Merge Healthcare, Inc.
iConnect Enterprise Archive. The firm name on the label is Merge Healthcare, Hartland, WI. iConnect Enterprise Archive is intended for use as a vendor neutral archive for storage and communications of medical images and data.
| Date | Jan 30, 2016 |
| Status | Terminated |
| Reason | Interventional Radiology (IR) images are stored as JPEG2k Lossless in Merge Enterprise Archive (EA) and are not displaying correctly in RadSuite. |
| Root cause | Software design |
| Action | The firm, Merge, sent an "URGENT: MEDICAL DEVICE RECALL" letters dated 1/29/2016 and 5/25/2016 via e-mail or via certified mail if they did not have an e-mail address for the customers. The letters described the product, problem and actions to be taken. The customers were instructed to do the following: you must discontinue using the affected version of RadSuite or iConnect Enterprise Archive with GE AW Workstation; use the workaround listed in the letter; complete and return the enclosed form no later than June 17, 2016; ensure that all users of the product are provided with the notification |
| Quantity | 96 sites potentially have the affected versions for both products |
| Distribution | US Distribution to: CO and OK. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.