ClearanceBrief
Recall Z-2757-2016 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Sorin Group USA, Inc.

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded

DateSep 8, 2016 (initiated Aug 12, 2016)
StatusTerminated
ReasonThe tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
Root causeProcess control
ActionLivaNova issued an Urgent Safety Alert letter via certified mail on August 12, 2016 to all affected customers. The letter instructed customers to check their inventory to determine if the Cannulae belongs to the product code and lots listed. If the customer does not have any of the products described in the communication, please complete the response form below and return it per the instructions indicated on the form; otherwise, please check the tip angle to determine if it is conforming to specifications. If specification is not met, please file a complaint through your normal process whethe
Quantity1205 units worldwide
DistributionWorldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030231CORONARY ARTERY PERFUSION CANNULA WITH SELF-INFLATING BALLOON
Sorin Group Italia S.R.L., Apr 18, 2003
K972503CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE
Sorin Group Italia S.R.L., Aug 26, 1997