Recall Z-2782-2016 ยท PCL Enzyme Linked Immunoabsorbent Assay, Rubeola Igm

Clark Laboratories, Inc. (dba,Trinity Biotech USA)

Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96 Determinations, IVD

DateSep 12, 2016 (initiated Jul 11, 2016)
StatusTerminated
ReasonThe product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive control to be invalid (running high out of range).
Root causeNonconforming Material/Component
ActionRecall notification letters (dated 7/11/2016) were sent by FedEx on 7/11/2016. Customers were instructed to discontinue use of any remaining inventory (discard on site if any remaining). Customers were also advised that if the test resulted in a valid run, such that all Controls and Calibrator met the Quality Control Criteria as stated in the package insert, the patient results can be reported and a re-evaluation of patient results is not necessary. For questions, please contact Trinity Biotech's Help Desk at 1-800-325-3424 Option #3.
QuantityUS: 6 kits, Foreign: 50 kits
DistributionDistribution to US nationwide including California, and China (Hong Kong).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140455CAPTIA MEASLES IGM
Trinity Biotech USA, May 22, 2014