Recall Z-2681-2016 ยท LLZ System, Image Processing, Radiological

Philips Electronics North America Corporation

PCR Compano Basic Reader using 100-127V as mains voltage. Model numbers: 732027, 732028, and 732040. Radiological image processing system (Computed Radiography).

DateAug 12, 2016
StatusTerminated
ReasonA defect component of the Compano reader power supply can cause the power supply to burn and some smoke to leave the housing.
Root causeComponent design/selection
ActionPhilips will send a Field Safety Notice to all affected customers on August 12, 2016. The notice informs consignees of the issue and informs them that Philips will replace the power supply unit of the PCR Compano Basic reader system of all affected units free of charge (FCO73200037).
Quantity388
DistributionWorldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OR, PA, TX, VT, WA, WI, WV, and WY; and, the countries of Canada, Brazil, Germany, Iraq, Japan, Korea, Kuwait,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964124PHILIPS COMPUTED RADIOGRAPHY
Philips Medical Systems, Inc., Dec 18, 1996