Recall Z-2627-2016 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.

DateAug 23, 2016 (initiated Jan 31, 2016)
StatusTerminated
ReasonPotential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than the first viewport.
Root causeSoftware design
ActionMerge Healthcare sent an Urgent Medical Device Recall letter dated January 28, 2016, and via email on January 31, 2013, to all affected customers. the letter identified the product the problem and the action needed to be taken by the customer. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached for
Quantity25 sites potentially have the affected versions
DistributionDistribution was made to AL, AZ, CA, IL, MD, MA, MI, MO, NV, NJ, NC, OH, PA, TN, and VA. Government distribution was made to OK. There was no foreign/military distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053281EVMS (ENTERPRISE VISUAL MEDICAL SYSTEM SOFTWARE)
Emageon, Inc., Dec 7, 2005